description
Includes the following units:
• fluorometric detector,
• Injector
• pump
• MultiChrom software
• set of columns and pre-columns
• NF-13 sample filtration kit
Specifications:
FLUOROMETRIC DETECTOR
Excitation wavelength range: 190 - 360 nm
Increments of setting wavelength: 2 nm
Fluorescence wavelength ranges: 380-600 nm; 420-600 nm;
480-600 nm;
Operational volume of the cuvette: 6 mm3
Detection limit: 5х10-9 g/cm3 of anthracene in acetonitrile
PUMP
Syringe volume: 5 ml
Maximum operational pressure: 11 MPa
Eluent flow rate range: 5-999 µl/min
Increments of setting flow rate: 1 µl
INJECTOR
Maximum operational pressure: 11 MPa
Replaceable loops: 2, 5, 10, 20, 50, 100 μl
Categories: | |